Health product recall

XN-L CHECK (2019-07-09)

Starting date:
July 9, 2019
Posting date:
August 2, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-70575



Last updated: 2019-08-02

Affected Products

XN-L CHECK

Reason

Sysmex America, Inc. (Sai) recognized a trend with complaints from XN-L-series customers reporting low recovery of the Red Blood Cell Distribution Width  (RDW) RDWCV and RDW-SD parameters on one or more of the three levels of the XN-L check hematology control lot 9158

Affected products

XN-L CHECK

Lot or serial number

9158

Model or catalog number

213569

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

651-0073

JAPAN